- Производитель:
- CAT M. Zipperer
The system consists of an ergonomic pump housing and a detachable programming and operating unit with key-pad and alphanumeric display. All relevant parameters, e.g. volume, flow-rate and flow direction, are set via the key-pad on the control unit, which can be used remotely from the pump. The key-pad has 3 function- and 2 hot-keys, plus an analogue encoder wheel for fast data entry, enabling intuitive adjustment of the set values. The back-lit LCD shows both set and actual volumes, flow-rate, and other functions.
A choice of units of volume, flow-rate and specific weight can be pre-programmed, e.g. µl, ml, mg, g, µl/s, ml/m. The PCON C controller has a RS232 interface as standard (1200-4800,8,N,1) and the analogue and digital I/O ports provide full flexibility.
Connection to the analogue input allows the pre-programmed flow-rate to be overridden by a voltage or current signal. The analogue output (0-10 V and 0-24 mA) can then be used to record and read out the actual flow-rate.
The pump heads are encased in robust stainless steel bodies and are available in 20 µl, 200 µl, 400µl and 1000 µl stroke volumes. They are also available with or without piston rinse port. The models "VCS" have a rinse port integrated into the ceramic cylinder, no piston sealing is necessary.
All wetted parts are made of ceramic (99,7% Al2O3), PTFE or ETFE
Pumps with built-in rinse pump for piston rinsing are available.
The pumps have incorporated software programs for many applications:
- STANDARD DISPENSE - dosing of a preset volume
- DILUTOR - diluting of a solution
- PIPETTE - simulation of a pipette
- DELAYED DISPENSE - an electronic dropping funnel
- precise dosing of a set volume
- manual dispensing
- time-controlled dispensing
- titrating
- diluting
- regulated drop-dispensing
- sampling with and without an air-segment
Technical data:
Model HPLH | 20 V | 20 VCS | 200 V | 200 VCS | 300 VCS | 1000 V | 1000 VS |
Piston diameter | 4 mm | 4 mm | 6 mm | 6 mm | 8 mm | 10 mm | 10 mm |
maximum flow rate | 10 ml/min | 10 ml/min | 100 ml/min | 100 ml/min | 200 ml/min | 400 ml/min | 400 ml/min |
Minimum step volume | 1 µl | 1 µl | 10 µl | 10 µl | 20 µl | 50 µl | 50 µl |
Minimum continuous Flow rate | 11.5 µl/min | 11.5 µl/min | 0.115 ml/min | 0.115 ml/min | 0.230 ml/min | 0.575 ml/min | 0.575 ml/min |
maximum counter pressure * | 5 bar | 5 bar | 5 bar | 5 bar | 3 bar | 1 bar | 1 bar |
Piston rinse port | yes | ||||||
rinse port integrated into ceramic | yes | yes | yes | ||||
Adjustment of Flow rate | digital | digital | digital | digital | digital | digital | digital |
Adjustment of Volume | digital | digital | digital | digital | digital | digital | digital |
Accuracy EV % | <= 1% | <= 1% | <= 1% | <= 1% | <= 1% | <= 1% | <= 1% |
Precision CV % | <= 0,5% | <= 0,5% | <= 0,5% | <= 0,5% | <= 0,5% | <= 0,5% | <= 0,5% |
TTL inputs | yes | yes | yes | yes | yes | yes | yes |
analogue current input 0-24 mA, adjustable | yes | yes | yes | yes | yes | yes | yes |
analogue voltage input 0-10 V, adjustable | yes | yes | yes | yes | yes | yes | yes |
analogue current output 0-24 mA | yes | yes | yes | yes | yes | yes | yes |
analogue voltage output 0-10 V | yes | yes | yes | yes | yes | yes | yes |
serial interface RS 232 | yes | yes | yes | yes | yes | yes | yes |
Time programming | yes | yes | yes | yes | yes | yes | yes |
Minimum delay time in delay mode | 1 sec | 1 sec | 1 sec | 1 sec | 1 sec | 1 sec | 1 sec |
Maximum delay time in delay mode | 100 h | 100 h | 100 h | 100 h | 100 h | 100 h | 100 h |
minimum flow rate in delay mode | 1 µl/100h | 1 µl/100h | 10 µl/100h | 10 µl/100h | 20 µl/100h | 50 µl/100h | 50 µl/100h |
Power supply | 24 V DC | 24 V DC | 24 V DC | 24 V DC | 24 V DC | 24 V DC | 24 V DC |
Dimensions W x H x D (mm) | 145x 170x 145 | 150x 170x 145 | 145x 170x 145 | 156x 170x 145 | 150x 170x 145 | 106x 170x 145 | 106x 170x 145 |
Order number | 70940- 00 | 70947- 00 | 70950- 00 | 70957- 00 | 70967- 00 | 70970- 00 | 70975- 00 |
Order number for pump with integrated rinse pump | - | 61772- 00 | - | 61775- 00 | 61777- 00 | - | 61779- 00 |
Model HPLH | 20 V | 20 VCS | 200 V | 200 VCS | 300 VCS | 1000 V | 1000 VS |
* depends on viscosity of the media and flow-rate
Powered by a stabilised 24V power supply, to be ordered separately, part no. 60732-00
Получите коммерческое предложение в течение 1 часа
Менеджер подготовит коммерческое предложение и позвонит, если понадобится уточнить детали вашего заказа

С 2010 года мы поставляем оборудование с заводов Европы. Берем на себя все — от подбора оборудования до внедрения на предприятии

Все сотрудники имеют высшее образование, закончили ведущие химические вузы страны, такие как РХТУ им Менделеева.

У большинства компаний срок ожидания составляет 10-12 недель.

Оборудование хранится на сухом отапливаемом складе, где поддерживается ровная температура.

Работаем с PonyExpress и Деловыми линиями. Вы также можете выбрать свою транспортную компанию или забрать товар со склада в Москве.

В случае любых неполадок за свой счет выполним ремонт в сервисном центре или на заводе-изготовителе. Или бесплатно заменим прибор на новый.

Производим пуско-наладку оборудования, валидацию, обучение сотрудников. Если нужно, привлекаем инженеров с заводов- изготовителей.